Open system with user dened proles and calculation chemistries
Reagent/Sample Handling:
Reagent/Sample tray:
40 positions for reagents and 40 positions for samples in refrigerated compartment (4~15°C)
Reagent volume:
R1:
180~450l, step by 1 l
R2:
30~450l, step by 1 l
Sample volume:
3~45l, step by 0.5 l
Reagent/Sample probe:
Liquid level detection, collision protection and inventory checking
Probe cleaning:
Automatic washing for both interior and exterior Carry-over < 0.1%
Automatic sample dilution:
Pre-dilution and post-dilution Dilution ratio up to 1: 150
Dilution vessel:
Disposable cuvette
Internal Bar Code Reader (optional):
Used for sample and reagent programming
Applicable to various bar code systems of Codabar, ITF (Interleaved Two of Five), code128, code39, UPC/EAN, Code93
Capable to communicate with LIS in bi-directional mode
ISE Module (optional):
Measuring K+, Na+, Cl−, LI+
Throughput:
Up to 225 tests per hour
Reaction System:
Reaction rotor:
Rotating tray, containing 80 cuvettes
Cuvette:
Optical length 5mm
Reaction volume:
180~500l
Reaction temperature:
37°C
Temperature uctuation:
±0.1°C
Mixing System:
Independent mixing bar
Optical System:
Light Source:
Halogen-tungsten lamp
Photometer:
Reversed optics, static ber spot photometry
Wavelength:
340nm、405nm、450nm、510nm、546nm、578nm、630nm、670nm
Absorption range:
0~4.0Abs (10mm conversion)
Resolution:
0.0001Abs
Control and Calibration:
Calibration mode:
Linear (one-point, two-point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline, Exponential, Polynomial, Parabola
Control software:
Westgard multi-rule, Cumulative sum check, Twin plot
Operation Unit:
Operation system:
Windows® XP Professional/Home SP2 or above Windows® VISTA Home/Business
Interface:
RS-232
Working Conditions:
Power Supply:
AC 200~240V, 50/60Hz, 1000W or
AC 100~130V, 50/60Hz, 1000W